ALAN's Electrosurgical Unit (Diathermy) is designed with latest miniature electronic circuit design technology. The advanced microcontroller based helps in achieving the desired tissue effects. It is small in size but Surprises on Performance.


SAFETY :
  •  Unit can work with input voltage of 230 VAC.
  •  High Frequency Leakage monitoring and controlling system.
  •  Equipotential port for ground connection.
  •  Built-in-line filter protects from voltage surges/bad line conditions.
  •  Cardiac floating output unit.
  •  Convection cooling system offers better unit's life.
  •  Handswitch and Footswitch short detection during self-test.
RELIABILITY :
  •  Power up Self-Test to ensure unit healthiness.
  •  Real time tissue impedance monitoring for power control.
  •  Multiple microcontroller supervision for RF power accuracy.
  •  Audio and Visual indication for easy interpretation.
  •  Different audible tones for activation, warning etc.
  •  Light indication for split or non-split patient plate connection.
  •  User settable maximum RF activation Timeout limit.

GENERAL FEATURES :
  •  Maximum output of 400 Watts
  •  Feather touch membrane keyboard for easy operation of Machine.
  •  Unit is well suited for Underwater, General, Gynaec & other surgeries
  •  Instant power adjustment against the tissue density & impedance variation.
  •  Monopolar output can be activated via handswitch & footswitch.
  •  Simultaneous coagulation facility in Monopolar
  •  Low Cut mode for reduced smoke in Laparoscopy & Open Surgeries
  •  Blend mode for cutting with haemostasis effect in special surgical applications.
  •  Desiccate mode for standard coagulation in Monopolar
  •  Additional Spray mode for non-contact superficial coagulation
  •  Micro coagulation in Bipolar mode for delicate tissues
  •  Light indication for split or non-split patient plate connection.
  •  Light indication for patient contact quality with return electrode.
  •  Built-in patient plate contact quality monitoring system for patient's safety
  •  PPCM deactivates output power if it senses improper contact between patient & return electrode to eliminate the risk of patient burn
  •  Unit is compatible with Single, Double & Three Pedal Footswitch


INPUT SUPPLY :

Input Supply : 230VAC +/- 20% 50 Hz

Loading Duty Factor : 10Sec (Active) / 30 Sec (Pause)

Output Control : Digital Display, Push Buttons in all Modes.


TRANSPORT AND STORAGE :

Ambient temperature range : -30 °C to +65 °C

Relative humidity : 0% to 90% non-condensing

Atmospheric pressure : 500 millibars to 1060 milibars

SAFETY STANDARD :

Basic Construction : Accordance with
IEC 60601 - 1,
IEC 60601 - 1 - 2 &
IEC 60601 - 2 - 2

Protection Class : 1

Unit Type : CF

All Output : Floating Output

Electrical Potential : Connecting Pin and Balancing Indicated by Symbol.